Models
 |
Temperature |
 |
Relative Humidity |
 |
3 Phase Current |
 |
4 Channel Pressure |
 |
Low Pressure Differential |
 |
3 Channel Thermocouple |
 |
8 Channel Thermocouple |
 |
Process Signal |
 |
8 Channel Temperature |
 |
Pulse |
 |
RTD Temperature |
ACR products
ACR JR
ACR OWL
Smart Reader
Smart Reader Plus
Nautilus
Power Watch
Home
Product Overview
Product Specifications
Software
Accessories
About AXIOM & ACR
Contact us
Site map
|
ACR SmartReader Plus Common Specifications
Multi-channel, Configurable, 12 Bit Data Loggers
Features
- Multiple channels
- Accessible remotely by modem
- Multiple loggers can be linked to form a network
- 32 K, 128 K or 1.5 MB memory
- 10-year battery life
Product Specifications (General)
SmartReader Plus Data Loggers share many common features. Please refer to individual model specifications for further specifications.
Size |
107 x 74 x 22 mm (4.2 x 2.9 x 0.9").
|
Weight |
110 grams (3.75 ounces). |
Case Materials |
NorylŪ Plastic.
|
Operating Limits |
-40oC to 70oC (-40oF to 158oF) and 0 to 95% Relative Humidity (non-condensing). |
Clock Accuracy |
+/-2 seconds per day. |
Battery |
3.6 volt Lithium, 1 Amp-Hour. |
Memory Sizes |
32 K (21,500 readings). 128 K (87,000 readings). 1.5 MB (1,000,000 readings). See specific models for availability. |
Sampling Methods |
1. Continuous (First-in, First-out - not available with sample rates faster than 8 seconds). 2. Stop when full (Fill-then-stop). |
Sampling Rates |
User selectable rates from 25 per second to one reading every 8 hours. (BP-101 battery pack or PS-201 Power Supply required for sample rates faster than 8 seconds. See Accessories section for details). |
Resolution |
12 bit (1 part in 4096). |
Power Consumption |
5 to 10 microamps (continuous). |
Battery Life |
10-years warrantied (under normal use). Factory replaceable. |
PC Requirements |
IBM PC or 100% compatible running MS Windows 3.1, '95, '98, or NT, with at least 2 MB RAM, 2 MB of hard drive disk space and one free serial port. |
Product Approvals |
Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets Class A radiated and conducted emission requirements of Section 2 of the Radio Interference Regulations of Communications Canada. IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/Burst and Radiated Emissions respectively for the Commission of the European Communities (CE). |
Resistance to X-Rays |
Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport X-Ray machine). Tested for protection against Gamma Ray (equivalent to 0.137 - 1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5".
|
Accuracy Curve |
Temperature Sensor Accuracy with ACR SmartReader Plus
|
Software Specifications |
TrendReader - SmartReader Plus Data Logger Software RealTime Graphing Software Talk-Back Software
|
Remote & Network communications |
ACR SmartReader Plus Networking and Remote Communications
|
Internal Temperature Sensor Specifications (Common on all models, except model 9)
Type |
Negative Temperature Coefficient Thermistor. 10,000 ohms @ 25oC (77oF).
|
Tolerance (sensor only) |
+/-0.2oC, over the range of 0 to 70oC (+/-0.3oF, over the range of 32 to 158oF).
|
Range |
-40 to 70oC (-40 to 158oF). |
Resolution |
0.03oC (0.05oF) at 25oC (77oF). Better than 0.07oC (0.12oF) between -25oC (-13oF) and 70oC (158oF). Better than 0.13oC (0.23oF) between -40oC (-40oF) and -25oC (-13oF). |
Home
| Product Overview | Product
Specifications | Software | Accessories
| About AXIOM
& ACR | Contact
us Site
map
You are visitor number