SmartReader Plus 2

Models
SmartReader Plus 1 Temperature
SmartReader Plus 2 Relative Humidity
SmartReader Plus 3 3 Phase Current
SmartReader Plus 4 4 Channel Pressure
SmartReader Plus 4 Low Pressure Differential
SmartReader Plus 5 3 Channel Thermocouple
SmartReader Plus 6 8 Channel Thermocouple
SmartReader Plus 7 Process Signal
SmartReader Plus 8 8 Channel Temperature
SmartReader Plus 9 Pulse
SmartReader Plus 10 RTD Temperature

Chinese version

ACR products
ACR JR
ACR OWL
Smart Reader
Smart Reader Plus
Nautilus
Power Watch

Home
Product Overview
Product Specifications
Software
Accessories
About AXIOM & ACR
Contact us
Site map

ACR SmartReader Plus 2 Specifications

Four Channel RH and Temperature Data Logger

SmartReader Plus 2 is a self-contained data logger that is able to record both temperature and relative humidity - right out of the box. The on-board RH sensor is temperature compensated permitting reliable, worry-free RH readings. For remote sensing, simply attach an external temperature or temperature/RH probe.

Features:

  • 12 bit resolution
  • 4 channels
  • 10-year battery life
  • Up to 128 KB memory
  • User-configurable
  • Accessible remotely by modem
  • Network capable
  • N.I.S.T. traceable measurements
  • Includes calibration certificate
SmartReader Plus 2 drawing    

Product Specifications
No. of channels Four. (One for the internal temperature sensor, one for the internal RH sensor, one for optional external RH and one for optional external temperature or resistance).
Relative Humidity Sensor Capacitive thin polymer film.
Range 0 to 95% (non-condensing).
Accuracy +/-3% RH from 10 to 90% (-20 to 40oC [-4 to 104oF]).
Resolution Better than 0.04% between 25 & 60% RH at 25oC (77oF).
Response Time Adequate ventilation reduces the response time which is approximately 5 minutes in still air.
Memory Size 32 KB, 128 KB.
Environmental Conditions Like all relative humidity sensors, when exposed to contaminants and/or extreme environmental conditions accuracy degradation could result. For maximum long-term stability, sensor should not be exposed or subjected to organic solvents, corrosive agents (strong acids, SO2, H2SO4, CI2, HCl, H2S, etc.), and strong bases (compounds with PH greater than 7). Dust settling on the sensor surface will not affect sensor performance except possibly to decrease the speed of response. For more information on the RH sensor's tolerance to chemicals, see Accessories section under Relative Humidity.
Accuracy Curve Temperature Sensor Accuracy with ACR SmartReader Plus
Common Specifications ACR SmartReader Plus Common Specifications
Software Specifications TrendReader - ACR SmartReader Plus Data Logger Software
RealTime Graphing Software
Talk-Back Software
Remote & Network communications ACR SmartReader Plus Networking and Remote Communications
Accessories ACR EH-020A Temperature & RH Probe
ACR ET Series Temperature Probe

Order Information
MODEL MEMORY CATALOG#
SRP-002 32 KB 01-0009
SRP-002-128K 128 KB 01-0113

Home | Product Overview | Product Specifications | Software | Accessories | About AXIOM & ACR | Contact us Site map

You are visitor number since 10th May 2000